1201.1 Scope

AMENDMENT
This section has been amended at the state or city level.
The provisions of this chapter shall govern the design and installation of piping and storage systems for nonflammable medical gas systems and nonmedical oxygen systems. All maintenance and operations of such systems shall be in accordance with the Fire Code.

[F] 1202.1 Nonflammable Medical Gases

AMENDMENT
This section has been amended at the state or city level.
Nonflammable medical gas systems, inhalation anesthetic systems and vacuum piping systems shall be designed and installed in accordance with NFPA 99C.
Exceptions:
  1. This section shall not apply to portable systems or cylinder storage.
  2. Vacuum system exhaust terminations shall comply with the Mechanical Code.
Nonmedical oxygen systems shall be designed and installed in accordance with NFPA 50 and NFPA 51.