// CODE SNIPPET
L 404.8 Steam Sterilizers
JUMP TO FULL CODE CHAPTER
Controls shall be installed to limit the discharge temperature of condensate or water from steam sterilizers to 140°F (60°C) or less. A venturi-type vacuum system shall not be utilized with vacuum sterilizers.
Related Code Sections
L 404.8 Sustainable Practices, Steam Sterilizers
Controls shall be installed to limit the discharge temperature of condensate or water from steam sterilizers to 140°F (60°C) or less. A venturi-type ...
Uniform Plumbing Code 2018 of Nevada > L Sustainable Practices > L 404.0 Occupancy Specific Water Efficiency Requirements > L 404.8 Steam Sterilizers
7.4.2 Space Ventilation-Hospital Spaces, Sterilization Rooms
Steam that escapes from a steam sterilizer shall be exhausted using an exhaust hood or other suitable means. Ethylene oxide that escapes from a gas ...
Guidelines for Design and Construction of Outpatient Facilities 2018 of Nevada > 7 Space Ventilation-Hospital Spaces > 7.4 Surgery Rooms > 7.4.2 Sterilization Rooms
7.4.2 Space Ventilation-Hospital Spaces, Sterilization Rooms
Steam that escapes from a steam sterilizer shall be exhausted using an exhaust hood or other suitable means. Ethylene oxide that escapes from a gas ...
Guidelines for Design and Construction of Hospitals 2018 of Nevada > 7 Space Ventilation-Hospital Spaces > 7.4 Surgery Rooms > 7.4.2 Sterilization Rooms
8.4.2 Space Ventilation-Outpatient Spaces, Sterilization Rooms
Steam that escapes from a steam sterilizer shall be exhausted using an exhaust hood or other suitable means. Ethylene oxide that escapes from a gas ...
Guidelines for Design and Construction of Outpatient Facilities 2018 of Nevada > 8 Space Ventilation-Outpatient Spaces > 8.4 Surgery Rooms > 8.4.2 Sterilization Rooms
8.4.2 Space Ventilation-Outpatient Spaces, Sterilization Rooms
Steam that escapes from a steam sterilizer shall be exhausted using an exhaust hood or other suitable means. Ethylene oxide that escapes from a gas ...
Guidelines for Design and Construction of Hospitals 2018 of Nevada > 8 Space Ventilation-Outpatient Spaces > 8.4 Surgery Rooms > 8.4.2 Sterilization Rooms