1224.34.1.2.3 Single-Photon Emission Computed Tomography (SPECT) Facilities
Go To Full Code Chapter
When provided shall include the following:
- Scanner room shall provide a minimum clearance of 4 feet (1218 mm) at each side and the foot of the table. Additional space shall be provided when SPECT is combined with CT, and include compliance with Section 1224.18.3 and shielding requirements in Section 1224.34.1.1.
- A control room shall be provided with a full view of the patient in the SPECT scanner.
- Imaging equipment room shall be provided in support of the equipment provided.
- Handwashing stations shall be provided throughout the SPECT suite at locations of patient contact and at locations where radiopharmaceutical materials are handled, prepared or disposed of.
Related Code Sections
1224.34.1.2.3 Interior Environment, Single-Photon Emission Computed Tomography (SPECT) Facilities
of the table. Additional space shall be provided when SPECT is combined with CT, and include compliance with Section 1224.18.3 and shielding requirements ...
1224.34.1.2.3 Interior Environment, Single-Photon Emission Computed Tomography (SPECT) Facilities
and the foot of the table. Control room. A control room shall be provided with a full direct view of the patient in the SPECT scanner. Computer equipment room ...
1224.34.1.2.3 Interior Environment, Single-Photon Emission Computed Tomography (SPECT) Facilities
used between
bays.
Control room. A control room shall be provided with a full view of the patient in the
SPECT scanner.
Computer equipment room ...
1224.34.1.2.2 Interior Environment, Positron Emission Tomography (PET)
1224.34.1.1.
Cyclotron room. Where radiopharmaceuticals
are prepared on-site, a cyclotron shall be provided.
Cyclotron facilities shall be located ...
1224.18.3 Interior Environment, Computerized Tomography (CT) Scanning
with the following:
A control room or alcove shall be provided that is
designed to accommodate the computer and
other controls for the equipment. A view ...