1224.15.3.5 Scrub Facilities
Go To Full Code Chapter
Scrub sinks shall be located outside of sterile areas. A minimum of two scrub sinks shall be provided in a surgical unit containing one operating room. Four scrub sinks shall be provided in surgical units containing two operating rooms. One additional scrub sink shall be provided for each additional operating room. Scrub sinks shall have water supply controls not requiring direct contact of the hands for operation.
Related Code Sections
1224.15.3.5 Interior Environment, Scrub Facilities
Scrub sinks shall be located outside of sterile areas. A minimum of two scrub sinks shall be provided in a surgical unit containing one operating ...
1224.15.3.5 Interior Environment, Scrub Facilities
Scrub sinks shall be located outside of sterile areas. A minimum of two scrub sinks shall be provided in a surgical unit containing one operating ...
1224.15.3.5 Interior Environment, Scrub Facilities
Scrub sinks shall be
located outside of sterile areas. A minimum of two
scrub sinks shall be provided in a surgical unit containing
one operating ...
1224.28.2.4 Interior Environment, Scrub Facilities
Scrub facilities with
hands-free operable controls shall be provided adjacent
to the entrance of procedure rooms ...
1224.28.4.5 Interior Environment, Scrub Facilities
Scrub sinks shall be
located outside of sterile areas. A minimum of one
scrub sink station shall be provided for each interventional
imaging ...