1324.5.1.1 Required Test Pressure
2021 Oregon Plumbing Specialty Code > 13 Health Care Facilities and Medical Gas and Medical Vacuum Systems > 1324.0 Performance Criteria and Testing Category 1 (Gases, Medical Surgical Vacuum) > 1324.5 Initial Pressure Tests — Medical Gas and Vacuum Systems > 1324.5.1 Shutoff Valve > 1324.5.1.1 Required Test Pressure
Go To Full Code Chapter
The test pressure for pressure gases and vacuum systems shall be 1.5 times the system operating pressure but not less than a gauge pressure of 150 psi (1034 kPa). The test pressure shall be maintained until each joint has been examined for leakage by means of a leak detectant that is safe for use with oxygen and does not contain ammonia. [NFPA 99:5.1.12.2.3.4, 5.1.12.2.3.5]
Related Code Sections
1324.5.1.1 Health Care Facilities and Medical Gas and Medical Vacuum Systems, Required Test Pressure
The test pressure for pressure gases and vacuum systems shall be 1.5 times the system operating pressure but not less than a gauge pressure of 150 ...
2021 Oregon Plumbing Specialty Code > 13 Health Care Facilities and Medical Gas and Medical Vacuum Systems > 1324.0 Performance Criteria and Testing Category 1 (Gases, Medical Surgical Vacuum) > 1324.5 Initial Pressure Tests — Medical Gas and Vacuum Systems > 1324.5.1 Shutoff Valve > 1324.5.1.1 Required Test Pressure
P2906.8 Water Supply and Distribution, Joint and Connection Tightness
Joints and connections in the plumbing system shall be gas tight and water tight for the intended use or required test pressure ...
P2503.9 Plumbing Administration, Test Gauges
Gauges used for testing shall be as follows: Tests requiring a pressure of 10 psi or less shall utilize a testing gauge having increments of 0.10 ...
1011.1 Boilers, Water Heaters and Pressure Vessels, Tests
with the requirements of the ASME Boiler and Pressure Vessel Code or the manufacturer's requirements, and such tests shall be approved. A copy of all test documents ...