1312.1.3.1 Remote Activated
JUMP TO FULL CODE CHAPTER
A remotely activated
shutoff valve at a supply manifold shall not
be used for emergency shutoff. For clinical
purposes, such a remote valve actuator shall not fail-closed
in the event of a loss of electric power. Where
remote actuators are the type that fail-open, it shall
be mandatory that cylinder shutoff valves be close
where the system is not in use. [NFPA 99:5.3.6.19.4]
Related Code Sections
1312.1.3.1 Health Care Facilities and Medical Gas and Medical Vacuum Systems, Remote Activated
A remotely activated
shutoff valve at a supply manifold shall not
be used for emergency shutoff. For clinical
purposes, such a remote valve actuator ...
California Plumbing Code 2016 > 13 Health Care Facilities and Medical Gas and Medical Vacuum Systems > 1312.0 Shutoff Valves > 1312.1 General > 1312.1.3 Emergency Shutoff Valves > 1312.1.3.1 Remote Activated
1406.2.1.5 Process Piping, Emergency Shutoff Valves
Readily
accessible manual or automatic remotely activated
fail-safe emergency shutoff valves shall be installed
on piping and tubing other than waste ...
California Mechanical Code 2016 > 14 Process Piping > 1406.0 Pipe, Tubing, and Fittings > 1406.2 Hazardous Process Piping (HPP) > 1406.2.1 Installation in Exit Corridors and Above Other Occupancies > 1406.2.1.5 Emergency Shutoff Valves
1406.2.1.5 Process Piping, Emergency Shutoff Valves
Readily
accessible manual or automatic remotely activated
fail-safe emergency shutoff valves shall be installed
on piping and tubing other than waste ...
California Mechanical Code 2019 > 14 Process Piping > 1406.0 Pipe, Tubing, and Fittings > 1406.2 Hazardous Process Piping (HPP) > 1406.2.1 Installation in Exit Corridors and Above Other Occupancies > 1406.2.1.5 Emergency Shutoff Valves
5003.2.2.1 Hazardous Materials — General Provisions, Design and Construction
and tubing shall be identified in accordance with ASME A13.1 to indicate the material conveyed. Manual valves or automatic remotely activated fail-safe ...
California Fire Code 2019 > 50 Hazardous Materials — General Provisions > 5003 General Requirements > 5003.2 Systems, Equipment and Processes > 5003.2.2 Piping, Tubing, Valves and Fittings > 5003.2.2.1 Design and Construction
5003.2.2.1 Hazardous Materials—General Provisions, Design and Construction
tank, cylinder or bulk source. Manual emergency shutoff valves and controls for remotely activated emergency shutoff valves shall be identified ...
California Fire Code 2016 > 50 Hazardous Materials—General Provisions > 5003 General Requirements > 5003.2 Systems, Equipment and Processes > 5003.2.2 Piping, Tubing, Valves and Fittings > 5003.2.2.1 Design and Construction