When provided, the mammography room shall comply with the following:
1224.18.6.1 Space Requirements
  1. Area. Mammography rooms shall be a minimum of 100 square feet (9.3 m2).
  2. Shielded alcove. Each x-ray room shall include a shielded control alcove. For mammography machines with built-in shielding for the operator, omission of the alcove shall be permitted when approved by the certified physicist.
1224.18.6.2 Handwashing Station
A handwashing station shall be provided within the procedure room.