1318.8.8 Cross-Connection Test - Category 3 Plastic Vacuum Piping Systems
JUMP TO FULL CODE CHAPTER
Initial cross-connection tests for plastic vacuum piping systems shall be conducted as follows:
- Tests shall be conducted to determine that no cross connections exist between any Category 3 plastic vacuum piping systems or Category 3 copper piping systems.
- The vacuum source shutoff valves for the vacuum piping systems shall remain closed during the tests, unless they are being used for the cross-connection test vacuum source.
- The cross-connection test vacuum shall be a minimum of 12 inch (305 mm) HgV.
- The source of test vacuum shall be connected only to the vacuum piping system being tested.
- The individual gas-powered device system gas outlets and vacuum system inlets shall be checked to determine that the test vacuum is only present at the vacuum piping system being tested.
- The cross-connection tests shall be repeated for each installed vacuum system with plastic piping.
- The proper labeling and identification of system outlets/inlets shall be confirmed during the tests. [NFPA 99:5.3.12.2.7]
Related Code Sections
1318.8.8 Health Care Facilities and Medical Gas and Medical Vacuum Systems, Cross-Connection Test - Category 3 Plastic Vacuum Piping Systems
that no cross connections exist between any Category 3 plastic vacuum piping systems or Category 3 copper piping systems. The vacuum source shutoff valves ...
Uniform Plumbing Code 2018 of Nevada > 13 Health Care Facilities and Medical Gas and Medical Vacuum Systems > 1318.0 Testing and Inspection > 1318.8 Cross-Connection Tests — Medical Gas and Vacuum Systems > 1318.8.8 Cross-Connection Test - Category 3 Plastic Vacuum Piping Systems
1308.6 Health Care Facilities and Medical Gas and Medical Vacuum Systems, Category 3 Systems
Category 3 systems shall comply with Section 1308.0 through Section 1309.0, except as follows:
Dental air and dental vacuum shall comply ...
Uniform Plumbing Code 2018 of Nevada > 13 Health Care Facilities and Medical Gas and Medical Vacuum Systems > 1308.0 Pipe Materials > 1308.6 Category 3 Systems
603.0 Water Supply and Distribution, Cross-Connection Control
Cross-connection control shall be provided in accordance with the provisions of this chapter. No person shall install a water-operated equipment ...
Uniform Plumbing Code 2018 of Nevada > 6 Water Supply and Distribution > 603.0 Cross-Connection Control
1308.0 Health Care Facilities and Medical Gas and Medical Vacuum Systems, Pipe Materials
]
Category 3 systems shall comply with Section 1308.0 through Section 1309.0, except as follows:
Dental air and dental vacuum shall comply ...
Uniform Plumbing Code 2018 of Nevada > 13 Health Care Facilities and Medical Gas and Medical Vacuum Systems > 1308.0 Pipe Materials
1318.8.2 Health Care Facilities and Medical Gas and Medical Vacuum Systems, Sources of Test Gas
Sources of test gas shall be disconnected from all piping systems except for the one system being tested. [NFPA 99:5.1.12.2.4.2] ...
Uniform Plumbing Code 2018 of Nevada > 13 Health Care Facilities and Medical Gas and Medical Vacuum Systems > 1318.0 Testing and Inspection > 1318.8 Cross-Connection Tests — Medical Gas and Vacuum Systems > 1318.8.2 Sources of Test Gas