1910.95(h) Audiometric Test Requirements
JUMP TO FULL CODE CHAPTER
Audiometric tests shall be pure tone, air conduction, hearing threshold examinations, with test frequencies including as a minimum 500, 1000, 2000, 3000, 4000, and 6000 Hz. Tests at each frequency shall be taken separately for each ear.
Audiometric tests shall be conducted with audiometers (including microprocessor audiometers) that meet the specifications of, and are maintained and used in accordance with, American National Standard Specification for Audiometers, S3.6-1969, which is incorporated by reference as specified in Sec. 1910.6.
Pulsed-tone and self-recording audiometers, if used, shall meet the requirements specified in Appendix C: "Audiometric Measuring Instruments."
Audiometric examinations shall be administered in a room meeting the requirements listed in Appendix D: "Audiometric Test Rooms."
The functional operation of the audiometer shall be checked before each day's use by testing a person with known, stable hearing thresholds, and by listening to the audiometer's output to make sure that the output is free from distorted or unwanted sounds. Deviations of 10 decibels or greater require an acoustic calibration.
An exhaustive calibration shall be performed at least every two years in accordance with sections 4.1.2; 4.1.3.; 4.1.4.3; 4.2; 4.4.1; 4.4.2; 4.4.3; and 4.5 of the American National Standard Specification for Audiometers, S3.6-1969. Test frequencies below 500 Hz and above 6000 Hz may be omitted from this calibration.
Audiometer calibration shall be checked acoustically at least annually in accordance with Appendix E: "Acoustic Calibration of Audiometers." Test frequencies below 500 Hz and above 6000 Hz may be omitted from this check. Deviations of 15 decibels or greater require an exhaustive calibration.
Related Code Sections
1910.95(h) Occupational Health and Environmental Control, Audiometric Test Requirements
the requirements listed in Appendix D: "Audiometric Test Rooms ...
OSHA 1910 General Industry > G Occupational Health and Environmental Control > 1910.95 Occupational Noise Exposure > 1910.95(h) Audiometric Test Requirements
1910.95(h)(5)(ii) Occupational Health and Environmental Control,
Audiometer calibration shall be checked acoustically at least annually in accordance with Appendix E: "Acoustic Calibration of Audiometers." Test ...
OSHA 1910 General Industry > G Occupational Health and Environmental Control > 1910.95 Occupational Noise Exposure > 1910.95(h) Audiometric Test Requirements > 1910.95(h)(5) Audiometer Calibration > 1910.95(h)(5)(ii)
1910.95(h)(5)(i) Occupational Health and Environmental Control,
The functional operation of the audiometer shall be checked before each day's use by testing a person with known, stable hearing thresholds ...
OSHA 1910 General Industry > G Occupational Health and Environmental Control > 1910.95 Occupational Noise Exposure > 1910.95(h) Audiometric Test Requirements > 1910.95(h)(5) Audiometer Calibration > 1910.95(h)(5)(i)
1910.95(h)(5)(iii) Occupational Health and Environmental Control,
; and 4.5 of the American National Standard Specification for Audiometers, S3.6-1969. Test frequencies below 500 Hz and above 6000 Hz may be omitted from ...
OSHA 1910 General Industry > G Occupational Health and Environmental Control > 1910.95 Occupational Noise Exposure > 1910.95(h) Audiometric Test Requirements > 1910.95(h)(5) Audiometer Calibration > 1910.95(h)(5)(iii)
1910.95(g)(7)(iii)(C) Occupational Health and Environmental Control,
Measurements of background sound pressure levels in the audiometric test room as required in Appendix D: Audiometric Test Rooms ...
OSHA 1910 General Industry > G Occupational Health and Environmental Control > 1910.95 Occupational Noise Exposure > 1910.95(g) Audiometric Testing Program > 1910.95(g)(7) Evaluation of Audiogram > 1910.95(g)(7)(iii) > 1910.95(g)(7)(iii)(C)